IEC 60601-1-Figure 6 Standard test finger
IEC61032 Figure 2,
Diameter: 12mm
Length:80mm
Insulated portion: Ø50mm, 20mm wide,100mm long
TESTING AND MEASURING EQUIPMENT/ALLOWED SUBCONTRACTING
IEC 60601-1:2005 + Am.1:2012
Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
Clause | Measurement/testing | Testing / measuring equipment / material needed | Subcontracting |
4.11 | Power input | Suitable devices for the voltage, current/power and frequency Supply: 1 phase and 3 phase variacs | R |
5.3 | Ambient temperature, humidity, atmospheric pressure | Suitable devices for recording ambient temperature, humidity, atmospheric pressure | R |
5.7 | Humidity preconditioning treatment | Environmental conditions: Climate chamber controlling temperature and humidity | R |
5.9.2 | Accessible parts | Force gauge (30 N), standard test finger (figure 6), straight unjointed test finger, test hook (figure 7) | R |
7.1.2 | Legibility of markings | Illuminance meter | R |
7.1.3 | Durability of markings | Distilled water, ethanol (96% pure), isopropyl alcohol , timer / stop watch | R |
8.4.2 | Accessible parts including applied parts | Oscilloscope, oscilloscope leads, suitable instruments for measuring voltage, current, capacitance, test pin (Figure 8), metal test rod (D = 4 mm, L = 100 mm), force gauge (10 N) | R |
8.4.3, 8.4.4 | Limitations voltage and energy | Suitable oscilloscope recorder /set-up & RCL meter | R |
8.5.5.1 | Defibrillation protection | 5 kV test circuit & oscilloscope interface circuit according to Figures 9 & 10, oscilloscope | S |
8.5.5.2 | Energy reduction test | The test circuit according to Figure 11, oscilloscope, oscilloscope leads | S |
8.6.4 | Impedance and current-carrying capability | Current source (25 A minimum, 50 or 60 Hz, 6 V maximum) | R |
8.7 | Leakage currents and patient auxiliary currents | Measuring device according to Figure 12, mains isolation transformers, variacs, voltmeter, millivoltmeter, aluminium foil, diverse circuits (fig. 13-20) | R |
8.8.3 | Dielectric strength | High voltage tester, isolating transformer for HV-tests (fig. 28), stop watch / timer | R |
8.8.4.1 a) | Ball pressure test | Test equipment according to IEC 60695-10-2 | R |
8.8.4.2 | Resistance to environmental stress | Apparatus for ageing rubber in oxygen | S |
8.9 | Creepage distances and air clearances | Oscilloscope, oscilloscope leads, callipers, micrometer, spacing gauges, force gauge (2 N & 30 N), standard test finger (figure 6) | R |
8.9.1.7 | Material groups classification | Test equipment according to IEC 60112 | S |
8.9.3.4 | Thermal cycling | Heating cabinet | R |
8.11.3.5 | Cord anchorage | Force gauge (at least 100 N), torque gauge (at least 0.35 Nm) | R |
8.11.3.6 | Cord guards | Weights, angle gauge, radius gauge | R |
9.4 | Instability hazards | 5° & 10° inclined planes or inclinometer or trigonometric calculation, force gauge (at least 220 N), 20 cm by 20 cm test surface, weights, test threshold (10 mm high and 80 mm wide), 7 cm strap, stop watch / timer | R |
9.5.2 | Cathode ray tubes | The relevant tests of IEC 60065, Clause 18. | S |
9.6.2.1 | Audible acoustic energy | A-weighted sound pressure level according to ISO 3746, ISO 9614-1 or IEC 61672-1 | S |
9.6.3 | Hand-transmitted vibration | Measurements are made in accordance with ISO 5349-1. | S |
9.7.5 | Pressure vessels | Hydraulic pressure test apparatus | S |
9.8 | Hazards associated with support systems | Weights or load cell, 0.1 m2 test surface, stop watch / timer, human body test mass (Figure 33) | R |
10.1 | X-radiation | Radiation meter | S |
10.3 | Microwave radiation | Radiation meter | S |
10.4 | Lasers | Test equipment according to IEC 60825-1 | S |
11.1 | Excessive temperatures | Temperature indicator/recorder suitable for this function and thermocouples, 4 wire resistance unit, test corner, variac | R |
11.2 | Fire prevention | Spark ignition test apparatus (Figure 34), oxygen gas analyzer | S |
11.3 | Constructional requirements for fire enclosures | FV tests specified in IEC 60695-11-10 | S |
11.6.2 | Overflow | 15° inclined plane or inclinometer or trigonometric calculation, stop watch / timer, high voltage tester | R |
11.6.3 | Spillage | Flask or graduated cylinder, stop watch / timer | R |
11.6.5 | Ingress of water or particulate matter | Classification tests of IEC 60529 | W |
11.6.6 | Cleaning and disinfection | Dielectric strength and leakage current tests as appropriate | R |
11.6.7 | Sterilization | Sterilisation to client specification | S |
13 | Hazardous situations and fault conditions | stop watch / timer, voltmeter, ammeter, temperature indicator / recorder suitable for this function and thermocouples, 4 wire resistance unit, cheesecloth | R |
15.3 | Mechanical strength | Force gauge (250 N minimum), circular plane surface 30 mm in diameter, 500 g steel ball, 50 mm thick hardwood board (hardwood > 600 kg/m3), 40 mm step, hardwood doorframe (40 mm2), circulating air oven | R |
15.4.2 | Temperature and overload control devices | Positive temperature coefficient devices (PTC’s) with IEC 60730-1: 1999, clauses 15, 17, J.15 and J.17 | S |
15.4.3.4 | Primary Lithium batteries | Performance of the tests identified in IEC 60086-4 | S |
15.4.3.4 | Secondary Lithium batteries | Performance of the tests identified in IEC 62133 | S |
15.4.6 | Actuating parts of controls | Force gauge (at least 100 N), torque gauge (at least 6 Nm), stop watch / timer | R |
15.4.7 | Cord-connected hand-held and foot-operated control devices | Force gauge (minimum1 350 N), 30 mm diameter test tool, stop watch / timer | R |
15.4.7.3 | Entry of liquids | Classification tests of IEC 60529 | W |
15.5.1.1 | Transformers | Winding tester for transformers, temperature indicator/recorder suitable for this function and thermocouples variac, loads | R |
15.5.2 | Dielectric strength | 5x voltage / 5x frequency dielectric strength test fixture, stop watch / timer | R |
16.6 | Leakage currents | As in sub-clause 8.7 | R |
A 10.4 | Light emitting diodes (LEDs) | As specified in IEC 62471 | S |
G | Protection against hazards of ignition of flammable anaesthetic mixtures | 4 mm & 12 mm diameter test rods, timer / stop watch, gasket test B-b of IEC 60068-2-2, force gauge (minimum 100 N), gas pressure meter (0 to 400 Pa range), voltmeter, ammeter, ohmmeter, meter for capacitance and inductance, oxygen mixture ( volume percentage 12,2 % ± 0,4 %),test apparatus for flammable mixtures (Figure G.7), | S |
G.4.3 | Prevention of electrostatic charges | antistatic material testing according to ISO 2882 | S |
L | Insulated winding wires for use without interleaved insulation | Sample is preparation according to IEC 60851-5:1996 & dielectric according to 60601-1; flexibility and adherence test 8 of IEC 60851-3:1996; mandrels of diameters according to Table L.1; heat shock test 9 of IEC 60851-6:1996; 2 mm diameter shot of stainless steel, nickel or nickel plated iron. | S |
KingPo Technology Development Limited
Kingpo Technology Development Limited is a pioneer and leader of global safety testing equipment manufacturer.We offer one-stop service for worldwide labs from R&D,testing,production to after-sales service.We dedicate ourselves in achieving optimal production quality for our customers,reduce cost and enhance enterprise’s international competitiveness.
Kingpo is accredited by International Electrotechnical Commission(IEC) and becomes a nominated supplier and is officially issued a CBTL certificate.We are now among the world’s leading laboratory testing equipment manufacturer to produce the highest level and more standard testing equipment for global Electrotechnics labs becoming nominated supplier of NEMKO,UL,PHILIPS,CTK,CVC(Vkan Certification & Testing Co.,Ltd),Dongguan Quality Testing,China Household Electric Appliance Research Institute,Russia National Standard Institute,Belarus National Standard Institute,Korea National Standard Institute,etc.Lamp cap gauges and testing equipments produced by Kingpo,which are calibrated by CNAS and ilac-MRA ,have a high international reputation.
Kingpo aims at manufacturing world class testing equipment for Chinese labs,establishing an independent R&D center and introducing foreign state-of-the-art technology and international high-tech talents.Adopting a creative design concept,taking sensor data acquisition as forerunner,computer control technology as lead,imported transmission device with high accuracy and low noises,we have developed more than one hundred and sixty products.
Lab software Lims system
Material combustibility testing equipment
ELectric tool testing equipment
Plug and socket testing equipment
Switch service life testing equipment
Electrical accessories testing equipment
Luminaire testing equipment
Household appliances testing equipment
Anti water spray testing equipment
Wire and cable testing equipment
IT product testing equipment
Battery testing equipment
Kingpo’s domestic customers include Entry-Exit Inspection and Quarantine Bureau, Bureau of Technical Supervision,Product Quality Supervision and Inspection Institute,UL Lab(Underwriter Laboratories),TUV Rheinland,SGS and other electronic manufacturer.Our products are exported to international market and have also established distributorships in HK,Vietnam,Singapore,Thailand,Iran,India,
Africa,New Zealand,Australia ,Germany and so on.
Service is an essential factor to improve enterprise’s competitiveness.Therefore,Kingpo has always made every effort in improve service level and service value.Before sale,we have professional engineering team to provide customers with comprehensive consultant and customized service.After sale,we have after-sale team to offer customer timely and satisfying technical support and product maintenance.
By means of excellent product competitiveness and complete service,our company has won numerous customers high praise and sincere trust.In complying with technical improvement and increased efficiency and quality assurance of these products ,we insist on providing high end products to market and dedicating to being a leader in testing equipment industry.